4.13. Two-lead Failover

Most resistive temperature sensors are wired to the Model 346 in a four-lead configuration so that the voltage measurement leads are independent of the current source leads. This eliminates the contribution of lead resistance to the sensor reading. If one of the four leads breaks, the four-lead measurement becomes invalid, and the input starts reporting non-deterministic readings that may be inadvertently used for improper temperature control.

The two-lead failover feature allows the Model 346 to continue providing usable readings on a resistive input after a one or both leads break on either the current, or the voltage pair. When failover engages, the Model 346 reroutes the affected lead pair, estimates the lead resistance contribution from the most recent valid four-lead reading, and applies that estimate as a compensation offset to the remaining two-lead measurement. Readings produced in this state are flagged in the questionable status register with the broken lead bit (bit 1) so that the user can distinguish them from four-lead readings.

Note

Two-lead failover is a system wide feature that can be enabled from the estimated readings card in the settings page; the maximum lead resistance can also be configured on the same page. Two-lead failover is only available on four-lead inputs and will not function on thermocouple inputs. Two-lead mode can also be enabled manually on a per-input basis by entering a fixed lead resistance value using INTL.

Interface Command: INTL

Theory

While the input is operating normally with all four leads intact, the Model 346 tracks the last non-questionable four-lead reading and uses it as a reference point if a lead break occurs later in operation. When broken lead detection determines that a lead has broken, the internal lead routing is then swapped so that the surviving lead pair carries both the source current and the voltage measurement. At this point, the raw two-lead measurement includes both the sensor resistance, the resistance of the leads, and an internal instrument resistance which needs to be compensated for.

For more information on broken lead detection, see Broken Lead Detection.

If Two-lead failover is enabled, the difference between the new two-lead reading and the saved four-lead reference is taken as the estimated lead resistance, and this estimate is subtracted from each subsequent reading so that the reported value continues to track the sensor. The two-lead resistance estimate is held constant until the input is disconnected (both lead pairs disconnect) or until the input broken-lead detection is reset. In addition to the lead resistance present in the setup between the instrument and the sensor, there is a nominal 200 ohms of internal resistance in series that is added to the measured lead resistance. This puts an upper limit on readings as the additional lead resistance is included in the measurement. For example, PTCs limited to 1 kOhm of resistance will effectively have a maximum measurable resistance of 800 ohms due to the 200 ohms of internal series resistance. This may cause issues when estimating the lead resistance near the maximum measurable resistance. Scanner option cards have an additional nominal 250 ohms of internal series resistance, further reducing the maximum measurable resistance for scanner inputs.

Note

The lead resistance estimate is only as accurate as the last valid four-lead reading. Sensor drift or temperature changes that occur after the lead break but before the broken-lead detection is completed are not captured by the estimate. Lead resistance also changes with temperature, resulting in less accurate readings at temperatures farther from the break temperature.

Setup

Two-lead failover is a system-wide setting that applies to all eligible RTD/Diode inputs. To enable or disable the feature from the front panel, navigate to the Estimated Readings settings page and toggle the Two-lead failover option. The maximum acceptable lead resistance can also be configured on this page, and should be configured to be close to the expected lead resistance in the system. If this value is left unconfigured, breaks that occur while the temperature is changing rapidly may result in inaccurate lead resistance estimation rather than a failure to estimate.

../_images/two_lead_failover_setup.png

Figure 4.26 Two-lead failover in the Estimated Readings settings page.

Interface Commands: TLFAIL

Use

Once Two-lead failover and Broken lead detection are both enabled, no further user action is required during normal operation. If a lead break is detected, the affected input transitions to the failover state automatically and continues reporting a sensor reading.

While an input is operating in the failover state, its readings are flagged with the broken-lead bit (bit 1) in the questionable status register, and the estimated readings behavior is applied to remote interface queries. Users monitoring the input over the remote interface can use this flag to distinguish failover readings from normal four-lead readings.

If both lead pairs eventually break, the input is reported as disconnected and no reading is produced. Reconnecting either pair of leads will return the input to normal four-lead operation; the lead resistance estimate is cleared at this point. The input may also be returned to a non-failed state manually by issuing a broken lead reset.

Interface Commands: TLFAIL, TLFAIL?, BRKRST, INTL, INTL?

Manual two-lead mode

In addition to the automatic failover behavior, the Model 346 supports a manual two-lead mode for installations where the sensor is intentionally wired with only two leads. In this mode, the user provides a fixed lead resistance value (in ohms) that is subtracted from every reading on the specified input, along with the internal instrument resistance. When a non-zero lead resistance has been entered, the input is treated as a two-lead measurement regardless of the broken lead detection state. To turn off this feature, set the lead resistance value back to zero.

Note

Unlike in automatic failover, the leads must be wired to the voltage measurement pins, the current source pins are not used in manual two-lead mode.

Interface Command: INTL