4.3. VM-10

The VM-10 is a single-ended or differential voltage measurement amplifier that can operate up to 10 V.

../_images/vm_horizontal.jpg

The VM-10 provides:

  • Single-ended (A) or differential (A-B) voltage measurement

  • Configurable ranges from 10 V to 10 mV

  • Seamless range changes

  • Configurable AC input coupling

  • A configurable hardware filter

../_images/VM-10.png

Figure 4.5 Module equivalent diagram

The settings for the VM-10 can be configured on the page for the module:

../_images/VM_settings.png

Figure 4.6 VM-10 settings

Range

The range determines the largest voltage that can be measured by the module. In general, the lowest range that can be used will provide the best performance. See instrument specifications for the performance characteristics of each range. The list of range bandwidths can be located on the instrument by tapping the info circle on the range setting.

If auto is selected, the VM-10 will automatically change ranges based on the input signal. The auto ranging algorithm uses the maximum and minimum peak measurements to make decisions. The algorithm will observe the signal over the specified averaging time or time constant before making a decision. Therefore it will take longer to change ranges when the averaging time is longer, for example when there is a large number of NPLCs.

Available ranges are 10 V, 1 V, 100 mV, and 10 mV.

Note

The 10 V and 1 V ranges are not available when the input filter is enabled and filter optimization is set to highest reserve.

Input Configuration

The input configuration determines what voltage is measured by the VM-10.

  • A-B The VM-10 measures the difference between the voltage on input A and input B.

  • A The VM-10 measures the voltage between input A and measure ground.

  • Ground The input is internally connected to measure ground.

The indicator LEDs on the VM-10 will change to indicate what inputs are presently being used by the measurement.

Interface Command: SENSe#:CONFiguration

Input Coupling

If Coupling is set to AC, a high pass filter with a cut-off frequency of 0.16 Hz is engaged. This blocks DC signals from being amplified but may reduce measurement accuracy.

Interface Command: SENSe#:COUPling

Note

The VM-10’s input bias current will cause a small DC offset voltage when coupling is set to AC. AC coupling should not be used when DC measurements are needed.

Input Filter

The VM-10 has the capability to filter the input signal. Both a high pass and low pass filter are available.

This filter is a hardware filter located in the module. This is in addition to any digital filtering configured in the M81-SSM. This is useful for rejecting large interfering signals and preventing them from being amplified.

Filter Optimization

In lowest noise mode, some gain is placed before the filters. This configuration provides the lowest noise, but causes overloads if an interfering signal is too large.

In highest reserve mode, all gain will be placed after the filters. This configuration can tolerate the largest interference, but the noise is higher.

Note

The 10 V and 1 V measurement ranges are not available when the input filter is enabled and filter optimization is set to highest reserve.

Optimization

The VM-10 can apply gain before and/or after the filter. The gain allocation is controlled with the Filter optimization setting.

In Lowest noise mode, some gain is placed before the filters. This configuration provides the lowest noise, but causes overloads if an interfering signal is too large.

In Highest reserve mode, all gain will be placed after the filters. This configuration can tolerate the largest interference, but the noise is higher.

Corner Frequency

The corner frequency can be selected from the listed options for each filter. If the corner frequency is set to None, the filter is disabled.

Rolloff

Each filter can be configured for single pole (6 dB/oct) or two pole (12 dB/oct) operation.

Default Settings

The table below lists the settings of the VM-10 upon initial power on or after settings are reset. For more information about default settings see the Default settings section.

Setting

Default state

Mode

DC

Averaging time

1 NPLC

Range

Auto, 10 mV

Input configuration

A-B

Coupling

DC

Input filter

Disabled

Filter optimization

Lowest noise

Low pass corner frequency

None

Low pass rolloff

6 dB

Dark mode

Disabled