4.2. BCS-10

The BCS-10 is a unique type of current source called a balanced current source. It provides up to 100 mA of DC and AC output current.

../_images/bcs_horizontal.jpg

The BCS-10 provides:

  • A balanced current output

  • Configurable ranges from 100 mA to 10 nA

  • Common mode control

  • Configurable AC coupling

../_images/BCS-10.png

Figure 4.3 Module equivalent diagram

The settings for the BCS-10 can be configured on the page for the module:

../_images/BCS_settings.png

Figure 4.4 BCS-10 settings

Caution

If the BCS-10 is disconnected from a load, it will output the full compliance voltage of 20V. For sensitive devices, ensure that the source is disabled when connecting the module to the device. For devices that are sensitive to, or may be damaged by such voltages, the user must take measures to ensure that the source is disabled prior to disconnection from and/or re-connection to the device under test or add external voltage limiting devices if such operation cannot be ensured.

Range

The range determines the largest current that can be sourced by the module. In general, the lowest range that can be used will provide the best performance. See instrument specifications for the performance characteristics of each range.

If auto is selected, the M81-SSM will select the lowest current range that will support the amplitude and offset each time they are configured.

The available ranges are 100 mA, 10 mA, 1 mA, 100 µA, 10 µA, 1 µA, 100 nA, and 10 nA.

Coupling

If coupling is set to AC, signals below 1.6 Hz are blocked from the output. This minimizes offsets on the output if DC and low frequency signals are not needed. When coupling is set to DC, the BCS-10 has a low output impedance at low frequencies. DC coupling is needed when sourcing low frequencies or output shapes that have DC components such as square waves.

Interface Command: SOURce#:COUPling

CMR

The common mode reduction (CMR) servo circuit controls the common mode voltage on the differential output of the current source.

Should CMR be on or off?

The BCS-10 is unique in that both the I+ and I- outputs are high impedance. This means that both the I+ and I- leads can be at any voltage between +10 V and -10 V. Something needs to determine what this voltage will be. In many cases, there is no external low impedance path to measure common (for example in a 4-wire resistor measurement with the VM-10). In such cases, CMR should be enabled. This means that the BCS will determine the operating voltage. In other cases, the load of the BCS may be grounded externally. In such cases, CMR should be disabled.

If CMR is on, should I use internal or external CMR?

In most applications, internal CMR is suitable. However, external CMR can be used to reduce errors arising from common mode voltage. This is especially relevant for low resistance measurements. Consider a 4 wire resistor measurement like the following:

../_images/BCS_internal_CMR.png

Suppose the BCS-10 is supplying a 100 mA excitation current. This current through a 1 mΩ device under test will produce a differential voltage of 100 µV. However, with internal CMR mode, the BCS will produce an average voltage of 0 on its output terminals. In doing so, it will produce a common mode voltage of 100 mV because of the 1 Ω difference in the current leads. The VM-10 now has to measure a 100 µV signal in the presence of 100 mV of common mode. The VM-10 has the ability to reject at least 80 dB of common mode. Even with this rejection, the VM will indicate 110 µV, resulting in a 10% error. This error can be resolved by using external CMR, connected as follows:

../_images/BCS_external_CMR.png

In external CMR mode, the BCS-10 will adjust the I+ and I- outputs to produce 0 V at the CMR sense point, which has been connected to the VM-10 B input. Now the VM-10 is presented with only 100 µV of common mode voltage, significantly reducing the error.

Guard Drive

If Guards are enabled, the inner sheath is driven with the voltage present on the center conductor. The guard minimizes effective capacitance and leakage current. In either setting, the guard should not be connected to ground.

Interface Command: SOURce#:GUARd

Disable On Overload

If Disable on Overload is enabled, the BCS-10 firmware will automatically disable the output when an overload is detected. Note that this feature is implemented in firmware, and will not respond as instantaneous as a hardware implemented compliance would.

Interface Command: SOURce#:DOCompliance

High and Low Current Output Limits

The current output limits are software limits preventing the user from entering an output that could potentially damage the module’s load.

When the shape is not DC, the limit is applied to the sum of the offset and amplitude. Low current output limit is the minimum current that a user will be allowed to enter, while the high current output limit is the maximum current the user will be allowed to enter. Both current limits are bound between -100 mA and 100 mA with the expectation that the lower current output limit must be a lower value than the high current output limit.

Interface Command: SOURce#:CURRent:LIMit:HIGH Interface Command: SOURce#:CURRent:LIMit:LOW

Default Settings

The table below lists the settings of the BCS-10 upon initial power on or after settings are reset. For more information about default settings see the Default settings section.

Setting

Default state

Shape

Sine

Frequency

1 kHz

Amplitude

0 pA

Offset

0 pA

Ranging

Auto, 100 nA

Common mode reduction (CMR)

Enabled

CMR feedback source

Internal

Sync

Disabled

Sync source

S1 (or S2 if module is S1)

Sync phase shift

0 degrees

Guard drive

Disabled

Coupling

Auto, DC

Dark mode

Disabled

Disable on Overload

Disabled

High current output limit

100 mA

Low current output limit

-100 mA