7.5. Error Messages
The following are warning and error messages that may be displayed by the Model 346 during operation.
Message |
Description |
|---|---|
DISABLED |
Input is disabled. Refer to Input Setup section. |
NO CURVE |
Input has no curve assigned. Refer to Input Setup section. |
S.OVER |
Input is at or over full-scale sensor units. |
S.UNDER |
Input is at or under negative full-scale sensor units. |
T.OVER |
Input at or over the high end of the curve. |
T.UNDER |
Input at or under the low end of the curve. |
Input microprocessor communication stopped |
The main microprocessor has lost communication with the sensor input microprocessor. |
NOVRAM Corrupt |
Invalid data or contents in NOVRAM—when this message appears, instrument settings have been reset to default. |
All outputs disabled due Sensor X surpassing TLIMIT |
The temperature reading on a sensor input has exceeded the Temperature Limit setting. A detailed message will follow, which includes a reference to which sensor input’s temperature limit has been exceeded. |
Heater X disabled due to load out of range |
A short or open circuit condition has been observed on 1 of the heater outputs. A detailed message will follow, which includes a reference to which output caused the condition. The output will be turned off when this occurs. |
Output X set to closed loop due to zone error |
A zone was entered into where the assocated input did not have a curve attached. |
Closed loop control on Output X has been halted |
|
Output on group X has been halted. |
A heater group that is actively controlling has a new physical heater added to it. |